40 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
The instrument has booted in a safe mode which does not communicate
with any of the internal boards. Normal instrument operation is not
possible. The firmware upgrade utility should be run to fix any
incorrectly programmed boards.
-241 Hardware missing; No RF board was detected.
The instrument requires an RF board to successfullly operate. The
board may not be plugged in or may have an incorrectly programmed
FPGA.
-241 Hardware missing; No Source Module connected
Cannot turn on Source Module Mode if source module is not
connected. Check cable connection.
-241 Hardware missing; SIO address does not exist
The expected board is not installed. Report this error to service
office.
-240 Hardware error; BBG Arb Fifo has emptied.
The Baseband board BBG FIFO has emptied prematurely. The BBG FIFO
buffers digital data from SDRAM Waveform memory. Report this error
to service office.
-240 Hardware error; The BBG Arb has not armed within the expected time
period. For external source data (LVDS interface), please recheck all
cables and signals. For internal source data (waveform memory), then
report this error to your nearest sales and service office.