EasyManua.ls Logo

Keysight N5172B EXG - Page 41

Keysight N5172B EXG
155 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight X-Series Signal Generators Error Messages 41
Error Messages
List of Error Messages
This error should not appear with normal Arb operation. This is a
hard fault if it should occur. If this error is seen for external
source data from the LVDS interface, then the interface and its
associated timing and control is suspected.
-240 Hardware error; BBG Arb SDRAM initialization failure: {name}.
There is a problem initializing the BBG waveform memory (SDRAM).
Report this error to service office.
-240 Hardware error; BBG Arb SDRAM refresh failure.
The baseband board BBG waveform memory failed a refresh cycle.
Report this error to service office.
-240 Hardware error; BBG Arb waveform memory (SDRAM) cannot be
accessed.
There is a problem accessing the BBG waveform memory (SDRAM).
Report this error to service office.
-240 Hardware error; BBG Communication failure
Communication with the BBG FPGA failed; specifically, the SSIO bus
failed to respond. Cycle the instrument's power. Report this
failure to service office.
-240 Hardware error; BBG DMA error: {name}.
An internal error occurred during DMA transactions to the Baseband
assembly. Please report this error to a service office.