42 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
-240 Hardware error; BBG Feldspar ASIC has failed to initialize.
The BBG Assembly Feldspar Asic has failed to initialize properly.
The instrument will require power cycling. Report this error to
your nearest sales and service office.
-240 Hardware error; BBG RMS Calculation has failed due to a {name} error.
The BBG RMS calculation has failed due to a timeout or an overflow
error.
-240 Hardware error; BBG waveform memory transaction aborted.
BBG temporarily stalled. No data was lost. The operation may have
exceeded the computational bandwidth. To prevent this error, use a
less complex waveform or stop the BBG before performing other
tasks.
-233 Invalid version
A legal program data element was parsed but could not be
executed because the version of the data is incorrect to
the device. This particular error is used when file or block
data elements are recognized by the instrument, but cannot
be executed for reasons of version incompatibility. For
example, a non-supported file version or a non-supported
instrument version.
-230 Data corrupt or stale; EEPROM @ offset 0x%06X: wrote 0x%02hX, read
0x%02hX