Keysight X-Series Signal Generators Error Messages 43
Error Messages
List of Error Messages
After writing shadow RAM data to EEPROM, a memory comparison
detected a difference. The EEPROM may not be retaining data
properly. Report this error to service office.
-230 Data corrupt or stale; Either the file is not a bit file, or the
number of bits specified does not match the number of data bytes in
the file.
An error has occurred while downloading a user file to the Bit
memory catalog. Either the user file is not a bit file, or the
number of bits specified in the SCPI command does not match the
number of bytes in the file. Ensure that the bit specifier matches
the actual number of bits in the file.
-230 Data corrupt or stale; File '{name}' Header Parameter Out of Range or
Not Valid: {name}.
This error is reported when trying to apply the header parameters
to the instrument state. If one of the parameters does not contain
a valid value for that state, this error is reported. The first {name}
is the file that generated the error. The second {name} will describe
which header field contained the invalid value.
-230 Data corrupt or stale; File is unusable: {name}
The file ({name}) is not properly formed, so it cannot be used. This
can happen if a file is not of the type expected, or the data was
damaged prior to being placed in the instrument. Note that when
importing ascii txt or csv files, a mismatch of the decimal
separator setting may cause this error message to be reported.
-230 Data corrupt or stale; Header File is Not Found or Invalid: {name}.