150 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
The first {name} is the file that generated the error. The %g is the
value of the Sample Rate as specified in the file header. The first
{value} is the lower limit of the sample rate in sps. The second {value} is
the upper limit of the sample rate in Msps.
703 Feature not supported; RS232 flow control is not supported.
Command is not supported for RS232 communication.
703 Feature not supported; The selected feature requires an option not
present on this instrument. To enable this feature, install option
'{name}'.
703 Feature not supported; The selected waveform was generated using an
option not present on this instrument. '{name}' See 'View Header
Details' for more information.
Selecting a file that was generated by another instrument with an
option that is not on the current instrument will generate this
error. {name} is the filename. Use the 'View Header Details' softkey in
the 'Header Utilities' menu for more information about the
licensing needed by segments and sequences.
703 Feature not supported; {name}. The column will be ignored.
This is reported when an imported list column requires missing
options not installed on the instrument.
704 Span adjusted due to selected sweep time; Choose a wider span or
shorter sweep time to avoid this conflict.