Keysight X-Series Signal Generators Error Messages         151
Error Messages
List of Error Messages
            *** No documentation ***
800 Digital module configuration error; Could not configure FPGA(s) on the
     Digital module: {name}. Report this error to service office.
            This error is reported when the FPGAs on the digital module FPGAs
            failed to be configured. The digital module will not work properly
            in this state.
801 Digital module VCO unlock error; The VCO on the digital module is
     unlocked. Check to make sure the clock settings are correct and the
     correct clock reference is connected.
            The VCO on the digital module is unlocked. Check to make sure the
            clock settings are correct and the correct clock reference is
            connected.
802 Digital module overrange error; The output data is being clipped by the
     resampler. Reduce the scaling  under the data menu to correct this
     problem.
            This error is reported when the output of the resampler is being
            clipped.
803 Digital module input FIFO overflow error; There are more samples being
     produced than can be consumed at the current clock rate. Verify that
     the digital module clock is set up properly.
            This error is reported when the digital module clock setup is not
            synchronized with the rate the samples are entering the digital
            module. Verify that the input clock rate matches the specified
            clock rate under the clock setup menu.