Keysight X-Series Signal Generators Error Messages 153
Error Messages
List of Error Messages
manual for details on how to update the instrument FW.
This error is reported when instrument FW has an older version than
the SSIO interface FPGA.
807 Unexpected digital module interface FPGA; The interface FPGA on the
digital module (0x{code},0x{code}) is not compatible with the
instrument FW. The supported FPGA version and code revision is
(0x{code},0x{code}). To ensure proper operation, please refer to the
manual for details on how to update the module's interface FPGA.
This error is reported when an SSIO interface FPGA is found with a
valid, but older version of the SSIO interface FPGA.
808 Digital module unlock error; The output clock is not locked. The clock
output may not be valid. Please make sure all references and clocks
are connected and the module is set up properly.
This error occurs when the output clock modules are not locked.
This can occur when the externally supplied clock is not stable,
the module is not set up properly, or the internal clock is not
locked. If the error persists after fixing all external causes,
then the module may need to be reset by turning it off then on.
This will set up the clock modules properly.
809 Digital module missing clock; Please make sure the external clock is
connected and the clock settings are correct.
This error occurs when the output clock modules are not receiving a
clock. This usually happens when the external or Device clock
source is selected and there is no clock source connected to the
clock input. If the error persists after fixing all external