Keysight X-Series Signal Generators Error Messages                   63
Error Messages
List of Error Messages
-221 Settings conflict; Channel #{value} must be turned on before channel
     #{value} can be turned on. Channel #{value} has been turned off.
            Transport channels must be turned on sequentially. For example,
            channels 1,2 and 3 must be turned on before channels 4 through 6
            can be activated.
-221 Settings conflict; Clipping not allowed for CDMA2000 or 1xEV-DV
     reverse radio configurations 1 and 2. Clipping set to 100%.
            The clipping function is not available for the CDMA2000 or 1xEV-DV
            reverse link radio configurations 1 and 2. The value is
            automatically reset to 100% (no clipping).
-221 Settings conflict; Clock Source Synchronized is only allowed when
     Logic Type is LVDS and  Data Direction is Input. Clock Source set to
     Internal.
            This error is reported when the Clock Source is set to Synchronized 
            and then the Logic Type is changed from LVDS or the Data Direction
            is set to Output.
-221 Settings conflict; The command cannot be executed when no TDMA or 
Real
     Time Custom mode is on.
            The command (e.g. MEM:DATA:PRAM:BLOCK) can be executed only
            when a TDMA mode (or Real Time Custom) is not on. Turn the desired
            mode on and then resend the command.
-221 Settings conflict; Compressed Mode should be off when {name}.