Keysight X-Series Signal Generators Error Messages 65
Error Messages
List of Error Messages
The data rate is changed to the default rate for the new radio
configuration, which is 1X for F-FCH and 2X for F-SCH.
-221 Settings conflict; The current paging indicator (PI) is not valid for
2400 bps data rate. The PI has been changed to 0, the default.
QPCH channel with a data rate of 2400 bps only supports up to 191
paging indicators. The PI has been set to 0 (default value) to
remain valid for this data rate.
-221 Settings conflict; The current RC or data rate does not support turbo
coding. Turbo Coding has been turned off.
Either specification or hardware does not support turbo coding for
this data rate. Turbo coding is now turned off. If turbo coding is
desired, select a valid data rate and then turn turbo coding on.
For F-SCH channels, data rates below 14.4 kbps and the highest data
rates for each RC are not valid for turbo coding.
-221 Settings conflict; Data of DPDCH must not be TrCH when method of
active Compressed Mode is Higher Layer.
Transport channel coding is supported on compressed mode by SF/2,
but it is not supported on compressed mode by Higher Layer. Higher
Layer method supports physical channel coding only.
-221 Settings conflict; DCH data block is too large. Maximum size of Block
Set ( = Block Size * number of Blocks) is 200000. Reduce Block Size
or number of blocks.
The 3GPP standard defines the transport block set size as 0 to
200000