74 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
The minimum sweep time is dependant on the sweep span and start
freq.
-221 Settings conflict; Internal burst cannot be on with pulse modulation.
There is a hardware conflict between internal burst and pulse
modulation. The most recently requested modulation will be turned
on. The previous modulation will be turned off.
-221 Settings conflict; Invalid Channel Code on DPCH{value}. The channel
code has been changed from {value} to {value}
The Channel Code is limited by the Slot Format. Parameter 1 is
channel number, Parameter 2 and 3 are channelization codes.
-221 Settings conflict; Invalid Channel Code on OCNS{value}. The channel
code has been changed from {value} to {value}
The Channel Code is limited by the Symbol Rate. Parameter 1 is
channel number, Parameter 2 and 3 are channelization codes.
-221 Settings conflict; Invalid Channel code. Channel code has been changed
to the highest Channel code for this symbol rate
An attempt was made to input a channelization code beyond the
limitations imposed by the symbol rate. The channelization code is
reset to the highest allowable code for the present symbol rate
setting.
-221 Settings conflict; Invalid Channelization code setting. Change
Channelization code setting of physical channels to be orthogonal.