Keysight X-Series Signal Generators Error Messages 75
Error Messages
List of Error Messages
OVSF codes require that physical channel code settings be
orthogonal.
-221 Settings conflict; I/Q and Ramp Sweep cannot be on at the same time.
{name}
The Ramp Sweep and I/Q modulation cannot be used at the same time.
This error is reported when the user attempts to activate one
feature while the other one in conflict is activated. If Ramp Sweep
is turned on then the I/Q feature will be turned off. If I/Q is
turned on then Ramp Sweep will be changed to Step Sweep.
-221 Settings conflict; I/Q cannot be turned on in the lowest frequency
band or with DSIM output. I/Q has been turned off.
The lowest frequency band (9 kHz up to 5 MHz) does not have WB AM.
-221 Settings conflict; I/Q delay adjustment limits depend on the rate of
the active format. The delay has been limited to the current upper or
lower bound as appropriate.
I/Q delay adjustment limits depend on the rate of the active
format. The delay has been limited to the current upper or lower
bound as appropriate.
-221 Settings conflict; I/Q must be turned on before turning on Constant
Phase.
Constant Phase cannot work with out an I/Q signal to the modulator,
so it does not make sense to turn Constant Phase on without I/Q on.