76 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
-221 Settings conflict; IR buffer size is too small.
The 3GPP standard defines the transport block set size as 0 to
200000
-221 Settings conflict; Marker Beacon mode must be ON with Marker Beacon
Pulsed mode. Marker Beacon mode has been turned ON.
The Marker Beacon Pulsed mode is a specific condition of Marker
Beacon mode; therefore, Marker Beacon mode has been turned ON.
-221 Settings conflict; The Marker Beacon Pulsed mode is not compatible
with Pulse Modulation mode. Marker Beacon Pulsed mode has been turned
off.
The Marker Beacon Pulsed mode adjustes the pulse modulator;
therefore, Pulse Modulation and Marker Beacon Pulsed mode ON are
not compatible.
-221 Settings conflict; The maximum value of the Summing Ratio is dependent
on the Modulator Atten value. Summing ratio value has been modified to
reflect the decreased IQ Modulator Atten value.
The maximum value of the Summing Ratio has a dependency on the
Modulator Atten value. Summing Ratio Max Value = +50dB +Modulator
Atten Summing Ratio Min Value = -50dB -Modulator Atten
-221 Settings conflict; Multi-BBG Sync is not supported with the current
Modulation Mode ({name}) and has now been switched off.
Multi-BBG Synchronization is not supported in this baseband
Modulation Mode.