90 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
set up the proper channels in addition to the R-SCH and turn on
turbo coding.
-221 Settings conflict; RSCH1 is only defined for Radio Configuration 3 and
4. You must set one of these Radio Configurations before configuring
or using this code channel.
The command attempted to enter a value or setting for the R-SCH1 in
RC3 or RC4. If this is the correct R-SCH, change to the desired
radio configuration and resend the command. If not, select the
command for the desired R-SCH. Use the following list to determine
the correct R-SCH command to use: R-SCH: Used in RC1 and RC2
R-SCH1: Used in RC3 and RC4 for Channel 4 R-SCH2: Used in RC3 and
RC4 for Channel 2
-221 Settings conflict; RSCH2 is only defined for Radio Configuration 3 and
4. You must set one of these Radio Configurations before configuring
or using this code channel.
*** No documentation ***
-221 Settings conflict; The sample clock rate must be less than or equal to
%gMHz for the ArbModulation Filter feature to be on. The sample
clock rate was reduced.
%g is the maximum sample clock rate to for the Arb Modulation
Filter feature.
-221 Settings conflict; Selected digital modulation cannot be on with pulse
modulation.
There is a hardware conflict between the selected digital