-221 Settings conflict; Walsh conflict between: {name} and {name}
            There is a Walsh collision between two channels specified. Either
            one Walsh code is equal to the other, or it falls under the other
            Walsh's tree. A Walsh collision results in an invalid signal.
            Select a different Walsh code according to Walsh selection rule.
-221 Settings conflict; Walsh length invalid! Changed to default Walsh
     number
            The current Walsh length is invalid for the current data rate. For
            data rate 2X and above, doubling the rate will half the walsh
            space. The channel Walsh number has been changed to the highest
            valid Walsh code.
-221 Settings conflict; Waveforms are not allowed in STEP sweep mode. This
     setting was turned  off.
 
-221 Settings conflict; WB AM cannot be turned on in the lowest frequency
     band or with DSIM output. WB AM has been turned off.
            The lowest frequency band (9 kHz up to 5 MHz) does not have WB AM.
-221 Settings conflict; When the clock rate setting is below 10 MHz or
     above 200 MHz, only phase settings of 0 and 180 degrees are valid. The
     phase was changed to a valid setting.
 When the frequency is changed to below 10 MHz or above 200 MHz and
            the phase is set at 90 or 270 degrees, the phase value is reduced
            by 90 degrees. Only 0 and 180 degrees are valid in the frequency