98                               Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
            ranges below 10 MHz and above 200 MHz.
-221 Settings conflict; When the noise carrier bit rate or carrier
     bandwidth are adjusted, the valid range of the Eb/No changes. The
     Eb/No has been limited to the new valid range.
 
-221 Settings conflict; You may not set Forward Link values while in
     Reverse Link. Reset Link Direction to proceed.
            Values or settings for forward link were entered while in reverse
            link. If forward link is the desired link direction, set the ESG to 
            forward link first, and then re-enter the values or settings.
            However, if reverse link is the correct link direction, consult the 
            SCPI or key references for the correct values or settings to enter.
-221 Settings conflict; You may not set Reverse Link values while in
     Forward Link. Reset Link Direction to proceed.
            Values or settings for reverse link were entered while in forward
            link. If reverse link is the desired link direction, set the ESG to 
            reverse link first, and then re-enter the values or settings.
            However, if forward link is the correct link direction, consult the 
            SCPI or key references for the correct values or settings to enter.
-221 Settings conflict; You may only set up channels that are part of your
     current radio configuration.
            An attempt was made to enter a value or setting for a channel that
            is not supported in the current radio configuration. The R-FCH and
            R-SCH are supported in all the radio configurations for reverse