Installation Note N9030-90045 39
Option BBA, Analog Baseband I/Q Inputs Retrofit Kit
Final Installation
1. Refer to Figure 3. Carefully slide the instrument cover back onto the instrument from the rear of
the analyzer, making sure not to damage any internal cables. The seam on the cover should be on
the bottom of the instrument. Be sure the cover seats into the gasket groove in the Front Frame
Assembly.
2. Replace the four rear feet (4) to the rear of the instrument using the four screws (3) . Torque to
21 inch pounds.
3. Replace the strap handles (2) on both sides of the instrument using the four screws (1). Torque to
21 inch pounds.
4. Replace the four instrument bottom feet.
5. Replace the four key locks to the bottom feet.
Install Instrument Software
http://www.agilent.com/find/pxa_software
This will guarantee:
The analyzer has the latest instrument software version to support Option BBA.
FPGA updates are done to the BBIQ hardware.
The drivers are installed for the new hardware.