40 Installation Note N9030-90045
Option BBA, Analog Baseband I/Q Inputs Retrofit Kit
Verify the Installation
1. Turn on the analyzer and wait for the instrument to fully boot up.
2. Verify the spectrum analyzer application loads and sweeps as expected. Press
System, Show,
Hardware on the analyzer and verify that the Main board and the BBIQ Interface Board identify
themselves as:
•BBIQ
BBIQ Interface
3. Press
System, Show, System on the analyzer to display a list of all installed options. You should
see N9030A-BBA.
Before performing adjustments and performance verification tests using the N7814A Calibration and
Adjustment Software for Agilent RF and Microwave Instruments, press
Mode, I/Q Analyzer,
Input/Output, I/Q, I/Q, I/Q Cable Calibrate... on the analyzer and follow the on-screen directions.
If the newly installed hardware is functioning properly the cable cal should work fine. Press
Mode,
Spectrum Analyzer, Input/Output, RF Input.