3 Short-Range Comms & IoT Mode
3.7 ACP Measurement
already selected
Loading a trace from a file puts that trace in View regardless of the state it was in when it was saved;
as does being the target of a Copy or a participant in an Exchange
Trace Update State On/Off
Remote
Command
For Swept SA Measurement (in SA Mode):
:TRACe[1]|2|…|6:UPDate[:STATe] ON | OFF | 1 | 0
:TRACe[1]|2|…|6:UPDate[:STATe]?
For all other measurements:
:TRACe[1]|2|3:<meas>:UPDate[:STATe] ON | OFF | 1 | 0
:TRACe[1]|2|3:<meas>:UPDate[:STATe]?
where <meas> is the identifier for the current measurement
Example Make trace 2 inactive (stop updating):
:TRAC2:UPD 0
Couplings Whenever you set Update to ON for any trace, the Display is set to ON for that trace
Preset For Swept SA Measurement (in SA Mode):
1|0|0|0|0|0
ON for Trace 1; OFF for 2–6
For all other measurements:
1|0|0
ON for Trace 1; OFF for 2 &3
State Saved Saved in instrument state
Trace Display State On/Off
Remote
Command
For Swept SA Measurement (in SA Mode):
:TRACe[1]|2|…|6:DISPlay[:STATe] ON | OFF | 1 | 0
:TRACe[1]|2|…|6:DISPlay[:STATe]?
For all other measurements:
:TRACe[1]|2|3:<meas>:DISPlay[:STATe] ON | OFF | 1 | 0
:TRACe[1]|2|3:<meas>:DISPlay[:STATe]?
where <meas> is the identifier for the current measurement
Example Make trace 1 visible:
:TRAC2:DISP 1
Blank trace 3:
:TRAC3:DISP 3
Couplings Whenever you set Update to ON for any trace, the Display is set to ON for that trace
Short Range Comms & IoT Mode User's &Programmer's Reference 995