Radiated Emissions Measurement Examples
Making Radiated Emission Measurements
Running Frequency Scan
Step Action Notes
1 Clear any existing
signal list
a. Press Meas Setup, Signal
List, Delete Signals,
Delete All.
2 Turn on the DUT and
start frequency scan
a. Turn the DUT on.
b. Press Meas Setup, Scan
Sequence, Scan Only.
c. Press Sweep/Control,
Start.
Let the Receiver take a number of scans
before going to the next step.
3 Stop the scan a. Press Stop.