Keysight X-Series Signal Generators Error Messages 33
Error Messages
List of Error Messages
-259 LAN error; The desired hostname could not be registered through
NetBIOS.
Contact your network system administrator for assistance.
-259 LAN error; Domain names can only contain letters, numbers, hyphens '-'
and dots '.'; the first character must also be a letter or number.
-259 LAN error; Hostnames are limited to 15 characters and can only contain
letters, numbers and hyphens. The first character must be a letter and
the last character must be a letter or a number.
-259 LAN error; The instrument could not extend lease for a dynamic IP
address using DHCP.
An initial DHCP LAN configuration was obtained. However, when the
lease expired, the instrument was unable to extend it and as a
result lost the previously assigned IP address. Refer to the
internet DHCP protocol documentation for further information.
-259 LAN error; The instrument could not obtain a LAN configuration using
Auto-IP: manual configuration was used instead.
The instrument was configured to use Auto-IP to obtain the LAN
interface configuration. Possibly, the LAN cable is disconnected or
is broken.
-259 LAN error; The instrument could not obtain a LAN configuration using
DHCP or Auto-IP: manual configuration was used instead. This error
occurs if no DHCP servers are present or the DHCP server did not
provide a lease of at least 5 minutes.