12
8. Judgment of Measurement Result
Compare the RLU measured in 7.5 with the label display on the light source.
Note that the label display for Lumitester Smart and
Lumitester PD-20 / PD-30 are different.
<Caution>
When a light source is measured with Lumitester Smart
and Lumitester PD-20 / PD-30, the measured valuesdiffer.
Therefore, each labeled RLU for Smart and RLU for
PD-20 / PD-30 are displayed.
When measuring LuciPac series, the measurement
result is the same for Smart and PD-20 / PD-30.
Results
Treatment
When it has not improved
1
When RLU is less than
80% of a label display.
(When a RLU display is
1000: 800 or less)
There is a possibility
that the measurement
chamber of Lumitester
or the luminescence
part window of the light
source is dirty.
Clean it off.
Failure of Lumitester
or a recalibration may be
required.
Please contact
the dealer
or us.
2 When RLU is over
120% of
a label display
(When a RLU display is
1000: 1200 or more)
There is a possibility
that the measurement
chamber of Lumitester
is dirty.
Clean it off.
Failure of Lumitester
or a recalibration may be
required.
Please contact
the dealer
or us.
3 When the measured
value of a blank tube
over 10RLU.
There is a possibility
that the measurement
chamber of Lumitester
is dirty.
Clean it off.
Failure of Lumitester
or a recalibration may be
required.
Please contact
the dealer
or us.
When Lumitester doesnt display a normal value by 1 or 2, and when it displays a
normal value by 3: The possibility of failure of a light source is also suspected.
Please send to the dealer or us together in repair of Lumitester or the case
of recalibration.
RLU display
for Smart
RLU display