SPECIFICATIONS
KI 9600 Series Optical Power Meter
Detector
type
Response
nm
Damage
level
dBm
Calibration
nm
Power
range
dBm
Tone & mulit-
fiber ID
sensitivity
dBm
Midrange
linearity
1
dB
Calibration
Accuracy
2
%
Polarization
Sensitivity
dB
Total
Uncertainty
3,5
dB
Sensitivity
5
± 30 nm dB
InGaAs
600 ~
1700
+15
660, 850
1300, 1310, 1390, 1490,
1550, 1610, 1625
+5 ~ -60
-40
-50
0.02
2 %
(0.09 dB)
< 0.005
0.3
0.03
H3B
(InGaAs)
800 ~
1700
+27
4
850
1300, 1310, 1390, 1490,
1550, 1590, 1610, 1625
+24 ~ -40
-20
-30
0.02
2 %
(0.09dB)
< 0.005
0.3
0.03
H5 (InGaAs)
800 ~
1700
+25
850
1300, 1310, 1390, 1490,
1550, 1590, 1610, 1625
+15 ~ -50
-30
-40
0.02
2 %
(0.09dB)
< 0.005
0.3
0.03
Ge
600 ~
1650
+15
635,650,660,780,1610,
1625
850, 1300, 1310, 1390,
1490, 1550
+10 ~ -60
-40
-50
0.04
2 %
(0.09dB)
< 0.005
0.5
0.04
typical
typical
max
typical
Note 1: Midrange linearity excludes top 3 dB and bottom 10 dB of range.
Note 2: Calibration condition: non coherent light, -35 ± 5 dBm, 23 ± 1
°
C, ± 1 nm, 10 ± 3 nm FWHM, PC ceramic connector, 100 µm fiber.
Note 3: Includes contributions of: varying optical connector types, calibration uncertainty, linearity over temperature & range, and fiber core diameter up to 200 µm.
Note 4: H3B can sustain the damage level for 2 minutes.
Note 5: At calibration wavelengths in bold type.
KI 9000 UM-10