KI 9000 UM-14 46
QUICK REFERENCE GUIDE – KI 9800 Series Optical Light Source
• Press and hold [LEVEL], then press [λ] for 3
seconds to set output power to factory default.
Alternatively, press [POWER].
AutoTest:
• To start AutoTest, press [POWER] then press [AUTO].
To end, press [AUTO] again or turn the instrument off.
TamperLock Mode:
• While in TamperLock Mode, user is prevented from
changing wavelength or laser output power.
• Press [POWER], turn the emitter on, then press and
hold [λ] while pressing [LEVEL] for 5 seconds.
Display will show “codE’.
• Enter 6 key sequence using any keys except
[POWER]. Triple beep and “LOut” on display will
indicate completion. To cancel your entry before
completion, press [POWER].
• While TamperLock is active, “<<<” and “>>>” symbols
will be displayed.
• To clear TamperLock Mode, turn off instrument then
press [POWER] while holding down [λ] and
[LEVEL]. When display show “codE”, enter six key
sequence.
• If the correct sequence had been entered, “PASS” will
be displayed indicating that TamperLock had been
cleared.
• In case where entered sequence was not correct,
“FAIL x” will be displayed (where x is attempt count).
• TamperLock will be cleared after three unsuccessful
attempts.