EasyManua.ls Logo

Kirsch ESSENTIAL-460 - Temperature Setting and Warning Limits; Changing Temperature Setting; Changing Temperature Warning Limits

Default Icon
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 9
7. Modifying temperature setting and temperature warning limits:
The values for the operating temperature and temperature warning limits are set at the factory. You must not
change these values. The following table also contains these values. The refrigerating machine requires a certain
amount of time to cool down to the set temperature. You can read the course of the temperature on the digital
display. The operating temperature is automatically maintained irrespective of changing ambient temperatures. The
condition for this is that the ambient temperature is at least 3°C above the set interior temperature.
WARNING!
Improper modifications to the temperature setting and warning limits can lead to irreparable damage to the
goods to be cooled! If in doubt, contact our customer services!
Changing the temperature setting:
o By pressing button (9) setted interior temperature is shown in display (2).
o To change the interior temperature, hold down button (9) and push buttons (3) and (6) to set the
desired target temperature.
Changing the temperature warning limits:
o By pressing button (3) and (6) as the same time for approx. 1-2seconds, the parameter (A1) appears
in display (2). Then you can change upper temperature warning limit.
o After that press button (9) and target temperature is shown on display (2).
o To change warning limits hold button (9) and adjust setting by pressing button (3) or (6).
o Let go button (9) and setting are saved automatically.
o By pressing button (3) parameter (A2) is shown in display (2), to set lower temperature limits.
o To change warning limits hold button (9) and adjust setting by pressing button (3) or (6).
o Let go button (9) and setting are saved automatically.
Information
On all devices, set the upper temperature warning limit at least 3°C higher, and the lower temperature limit at least 3°C
lower than the set interior temperature.
MODEL
Operating
tempera-
ture
°C
Lower warning
limit
°C
Upper warning limit
°C
ESSENTIAL-100
5
1
10
ESSENTIAL-280
5
1
10
ESSENTIAL-460
5
1
10
ESSENTIAL-700
5
1
10

Related product manuals