K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual -16-
12. Power Switch: Controls power to the entire unit. To power on the instrument, first press the Power
Switch to ON then press the Power Button (Figure 1, Item 1) on the front on the analyzer.
13. AC Outlet: The analyzer has been designed with an integrated C13 line socket in order to easily
change plug types (115V or 230V). Internal voltage can be switch using the Voltage switch (Figure 2,
Item 19).
14. Cooling Fan: Integrated cooling fan is capable of operating at two speed for either standard or rapid
cooling. The user may change the fan speed using the touch screen control panel.
15. USB Port: For connection to Keyboard or Mouse.
16. RS232 Port: For connection to external PC
17. Ethernet Port: For connection to LIMS system.
18. CPU Port: For programming of internal CPU. FOR FACTORY USE ONLY
19. Voltage Switch: The user may easily switch the internal voltage compatibility by switching up or
down. The voltage type will be displayed on the switch accordingly.
20. Inert Gas Inlet: For connection of external inert gas source, typically Nitrogen. Analyzer must be
connected to gas source for the integrated Fire Extinguishing System to be utilized properly.
21. LPG or Natural Gas Inlet: For connection to Natural Gas Source or Bottled Gas (LPG - Propane or
Butane is typically used).
22. Barometric Pressure Sensor: Point where internal automatic barometric pressure adjustment
system senses atmospheric pressure. No user function.