K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual -7-
Once analysis is complete the results are
displayed to the screen and saved to the
instrument hard disk drive. The results may be
opened in a report generation program like MS
Word or Excel and sent to a local or network
printer. Once the results have been reviewed
and accepted by the operator the results may be
sent directly to a network folder for processing
and direct entry into LIMS.
Once analysis is complete the instrument
actuates the rapid cool-down system. The cool-
down system is capable of reducing the
temperature of the sample from 300 to 30°C in
about 8 minutes in a laboratory with ambient
conditions at 20°C. User can also manually
switch to standard air cooling. Cooling the
system will take a longer period of time however
fan noise will be reduced.
1.4 Instrument Specifications
Models:
K71000
Electrical
Requirements:
115/230VAC 50/60Hz
User selectable via switch
located at the back of the
analyzer
Temperature
Measuring Range:
Ambient to 405°C (761°F)
Flash Detection
System:
Differential
Thermocouple,
Ionization Detection
Stirring Speed:
0 to 300 RPM
Heating Rate:
In accordance to ASTM
D93 procedure A, B, C
Quick Test (≈12°C/min)
Ignition Source:
Gas and Electric Igniter
both equipped on cover
assembly:
- Software Selectable
- User-Friendly Manual
Switching
Mechanical Lift:
Three (3) position
mechanical lift system of
cover / motor assembly,
software selectable:
Open Clean Test
Cleaning:
Cover / Motor assembly
lift features convenient
cleaning position. At this
position the user can
easily remove the cover
assembly and clean with
appropriate solvent
between tests.
Ignition Test
Frequency:
User selectable on per
method basis
Cooling:
Integrated Dual Fan
System: First (1) directly
to cup, Second (2) to cool
environment around test
cup
Barometric
Pressure
Correction:
Automatic Barometric
Pressure Transducer for
Flash Temperature
Correction
Interface Port:
RS232:1
USB:2 (1Front, 1Back)
Ethernet:1 (LIMS)
Gas Supply /
Connection:
Ignition Gas: LP Gas
(Propane / Butane) or
Natural Gas.
Min. 3.45kPa (0.5psi).
Max. 6.90kPa (1.0psi)
Display:
8.4" LCD Touch Screen
Temperature
Sensor:
PT-100 in S/S sheath,
User
Programmable
Methods & Data
Storage:
Capable of storing
unlimited data runs and
test methods