KPS-TL10 • Digital earth resistance tester
2. SAFETY NOTES
- To avoid electrical shock, remove test leads before opening case.
- Do not exceed 200 V AC in 3 phase tester.
- Do not apply voltage across COM and OHM terminals.
WARNING
To avoid electrical shock or damage to the meter, do not get
water inside the case.
3. SPECIFICATIONS
Measurement system: Earth Resistance by constant current
inverter. 820Hz, 2mA approx.
Earth Voltage: 0-200 V AC, 40-500 Hz.
Earth Resistance: Range and resolution
0-20 ohm (0.01 ohm)
0-200 ohm (0.1 ohm)
0-2000 ohm (1 ohm)
Accuracy: Earth voltage +/- (1%rdg + 2dgt)
Earth resistance +/- (2%rdg +
2dgt)or+/-0.1Ω,whichisgrater
Safety standard: IEC-1010 (EN 61010), installation
category III
Low battery indication: “B” symbol appears on the display.
Data hold indication: “DH” symbol appears on the display.
Over range indication: “1” (MSD).
Timer (auto power off): About 3-6 minutes.
Display: LCD31⁄2digit(2000counts)
Power source: (R6P) 1.5Vx6 pieces dry batteries.
Dimensions: 163 (L) x 100 (W) x 50 (D) mm.
Weight: Approx. 800g. (battery included)
Accessories: Test leads (red – 15M, yellow – 10M, green –
5M), Auxiliary Earth Bars Heavy Case,
Instruction Manual.