Applications:
Measuring currents in DC operated electrical appliances, bias current of
transitors, IC's, etc.
Insert the red test lead into the + termin~1 and the black test lead
into the
COM terminal.
Set the range selector switch to the 250mA range position.
Turn off power to the circuit under test.
Connect the red test lead in series with the positive (+) side of the
circuit under test and the black test lead with the negative (-) side.
Turn on power to the circuit under test.
When the reading is below 25mA, set the range selector switch to
the lower range for more accurate reading and proceed with the
measurement.
6-7. Resistance Measurements
#
WARNING
Be sure to turn off power to the circuit under test before making
resistance measurements.
#
WARNING
Never apply voltage to the ohm ranges. The fuse could blow or the
instrument get damaged. Especially when a voltage higher than
250V is applied, the fuse may not protect the circuit.
#
CAUTION
A maximum voltage of 12V is present on the ×10kΩ range. It may
damage IC's for low voltage equipment.
Before making resistance measurements, carefully check the
withstand voltage of the circuit under test.