19
dAY
Select to download a number
of days ie 5. Touch the tick
icon/button and use the
arrows to select the number
of days required, then touch
the tick icon/button for the
download to commence.
LSt
Select to download all the
data since the last download.
Touch the tick icon/button to
select and for the download
to commence.
All
Select to download all the
data held by the controller.
Touch the tick icon/button to
select and for the download
to commence.
The controller will rst check the card then begin to download the data. You will see the
screens below showing the progress of the download and End when it is completed.
If you see a screen with Err, an error has occurred so try again. If you see Er1, it means
initialization has failed, please contact our service team on 01256 705 580 or service@
labcold.com, and if you see ER2 it means the wrong Micro SD card for that refrigerator
is being used. Try again with the correct Micro SD card.
Viewing the data
When you insert you insert the SD card in to your computer you will see csv folders
named in yymm format. They contain CSV les named in yymmdd format. Click on the
file to select the date you want to download or view.
The csv le will be a mixture of temperature readings against time and fridge operation
data such as whether the door was open etc and the time it happened. A zero in a
column means no incidence and a one in a column means there was an incidence so in
the example below, the compressor was on at 00:11 but o at 00:12.