11
Close Sash
When input configured as Close Sash is activated
Red LED on
Close Sash toggles on /off with display
Audible alarm sounds (after pre-set time)
Audible can be muted via Enter pushbutton -- this silences the alarm if configured.
Mains Fail (Optional)
When the input configured as Mains Fail is activated
Red LED on
Mains Fail is displayed
Audible alarm sounds
Audible can be muted via Enter pushbutton -- this silences the alarm if configured.
Reset when Mains power is provided to monitor
Temperature (Optional)
When input configured as Temperature is activated (and temp sensor connected)
Temperature and Airflow displayed Temp display can be disabled.
High Temp toggles with Airflow display - Red LED on (Constant) with Audible alarm
Low Temp toggles with Airflow display - Red LED on (Constant) with Audible alarm