1.4 Events / actions
Safe airflow
Meter reading above warning level ( e.g. > 90fpm )
Green LED on
Warning airflow
Meter reads between warning level and air fail level ( e.g. > 80fpm and < 90fpm )
Amber LED on
Low airflow
Meter reads below alarm level for longer than the warning to low air delay time
AIR FAIL toggles on / off with display
Red LED on ( Flashing )
Audible alarm sounds -- can be muted via Enter pushbutton
Low air relay operates ( if configured )
Reset : -- when airflow rises 4fpm above Low air level for longer than the low air to
warning air delay time the Low air alarm resets automatically
High airflow
If configured :-
High Air toggles on / off with display
Audible alarm sounds can be muted via Enter pushbutton)
High air relay operates ( if configured )
Audible alarm Mute
When the audible alarm is muted via the Enter button - an Icon ( horn with forward slash) is
shown on the display.
Sash High
When the input configured as Sash High is activated
Amber LED on
Sash High toggles on / off with velocity display