6.0 Parameter list / factory settings.
AFA10001/Mk2 Labconco
FILE NUMBER : 1.6
TYPE : AFA1000/1Mk2
CUSTOMER : Labconco
The AFA1000 monitor has been factory set to the listed parameter settings. To change
any parameter setting press and hold the enter button until the "Main Menu" screen is
displayed, select "Configure" to view the parameters and use the +/- and Enter buttons
to change the settings.
FOR FULL INSTALLATION / CALIBRATION / CONFIGURATION DETAILS PLEASE
REFER TO THECALIBRATION SECTION OF THE USER MANUAL.
FACTORY DEFAULTS
Menu
Units
Default Settings
Configuration Menu: -
Calibration Config
ON
DISPLAY UNITS
M/SEC
FPM
FPM
ON
LOW AIR ALARM
FPM
60
ON
LOW AIR CUTOFF
OFF
ON
ON
ON
CUTOFF VELOCITY
FPM
30
ON
WARNING AIR ALARM
FPM
63
ALARM TO WARNING AIR HYSTERESIS
FPM
3
ON
HIGH AIR ALARM
ENABLED
DISABLED
DISABLED
ON
HIGH AIR ALARM
FPM
ON
LOWER AIR SAMPLE FLUCTUATIONS
%
5
ON
HIGHER AIR SAMPLE FLUCTUATIONS
%
10
ON
LOWER/HIGHER AIR SAMPLE DIFFERENCE
FPM
30
ON
WARN TO ALARM AIR TIME
SECONDS
30
ON
ALARM TO WARN AIR TIME
SECONDS
1
ON
SHOW AIR FLOW
ENABLED
ENABLED
DISABLED
ON
SHOW TIME LINE
ENABLED
DISABLED = SHOW BAR GRAPH
DISABLED
ON
AUDIBLE ALARM
ENABLED
ENABLED
NOT ENABLED
ON
SENSOR DIFFERENCE
%
10
ON
SENSITIVITY
%
100
Configuration Menu