67
Wash 1, Steam, Rinses 1 & 2, Dry Time +/- 1 minute
Wash 2, Final Rinse Time after reaching SP +/- 1 minute
Conductivity varies between 100 - 1200
Confirm operation of low detergent and acid rinse monitor
Detergent Dispense 18 - 22 ml
Rinse Aid Dispense 5 - 7 ml
5 Observe Wash Action – Water hits top of tank from spindles
INSP/TESTS No. CONDITIONS AND REQUIREMENTS INST/LCC# RESULTS PASS
6 Run rinse only to wet surfaces. Visual
After wash pump starts, place a wet towel across the overflow contacts.
Wash pump should stop and the washer should drain. Display should
show ALARM OVERFLOW.
Hi-Pot
1 1000V for 1 minute. Hi Pot Tester ____________
Final
1 Check for Leaks Visual
2 Unit must be Clean and Dry Visual
3 All parts in place in good condition Visual
4 Verify that thermostat on back of tank is installed properly. Visual
5 Make sure drain hose is routed over top of wash pump. Visual
6 Perform FACTORY RESET Visual