5.3.9 - StreamStop
Not supported over UDP.
Command:
Byte
0
0xB0
1
0xB0
Response:
Byte
0
Checksum8
1
0xB1
2
Errorcode
3
0x00
5.3.10 - ReadMem
Reads 1 block (128 bytes) from the Control non-volatile memory. The Mem area is arranged in 16 blocks of 128 bytes each.
Blocks 0-7 are used by LabJack Corporation to store calibration data, and blocks 8-15 are available to the user.
Command:
Byte
0
Checksum8
1
0xF8
2
0x01
3
0x2A
4
Checksum16 (LSB)
5
Checksum16 (MSB)
6
0x00
7
BlockNum
Response:
Byte
0
Checksum8
1
0xF8
2
0x41
3
0x2A
4
Checksum16 (LSB)
5
Checksum16 (MSB)
6
0x00
7
BlockNum
8-135
Data
5.3.11 - WriteMem
Writes 1 block (128 bytes) to the Control non-volatile memory. The Mem area must be erased before writing. The Mem area is
arranged in 16 blocks of 128 bytes each. Blocks 0-7 are used by LabJack Corporation to store calibration data, and blocks 8-15
are available to the user.
65