8 – LABNET BIOPETTE PLUS PIPETTORS
2-1,000 μL pipettors
M
M
4
M
M
4
M
M
4
1. Tip ejector disassembly
Press the tip ejector pushbutton.
Turn the tip ejector bush 1/3 turn counter-clockwise.
Slide the ejector from the plastic arbor.
2. Tip ejector assembly
Press the tip ejector pushbutton.
Line the tip ejector up with the plastic arbor.
Turn the tip ejector bush clockwise until it has fully
latched.
3. Tip ejector adjustment
Remove the tip ejector pushbutton.
To increase the ejector length turn the plastic bush
counter-clockwise.
To decrease the ejector length turn the plastic bush
clockwise.