Instruction Manual – 5
5. Specications
The pipettor is a high quality instrument which oers
excellent accuracy and precision. The accuracy and precision
(repeatability) of the liquid volume depends on the quality
of pipet tips used. The values for accuracy and precision
shown in the table below were obtained using manufacturer
non-lter pipet tips. Those tips are recommended for use to
ensure compatibility, accuracy and precision when pipetting.
Nominal
Volume
(μL)
Volume
(μL)
Accuracy
(%)
Precision
(%)
Non-lter
Tips (μL)
Single-channel Pipettors
2
Min. 0.1 ±40.0 ≤12.0
10
0.2 ±12.0 ≤6.0
1 ±2.7 ≤1.3
Max. 2 ±1.5 0.7
10
Min. 0.5 ±4.0 ≤2.8
10
1.0 ±2.5 ≤1.8
5.0 ±1.0 0.6
Max. 10.0 ±0.5 ≤0.4
20
Min. 2 ±3.0 ≤1.5
20010 ±1.0 ≤0.5
Max. 20 ±0.8 0.3
100
Min. 10 ±1.6 ≤0.80
20050 ±0.8 ≤0.24
Max. 100 ±0.8 ≤0.20
200
Min. 20 ±1.2 ≤0.60
200100 ±0.8 ≤0.25
Max. 200 ±0.6 ≤0.20
1,000
Min. 100 ±1.6 ≤0.40
1,000500 ±0.7 ≤0.20
Max. 1,000 ±0.6 ≤0.15
5,000
Min. 1,000 ±0.6 ≤0.25
5,0002,500 ±0.6 ≤0.20
Max. 5,000 ±0.5 ≤0.15
10,000
Min. 1,000 ±2.5 0.6
10,0005,000 ±0.8 0.3
Max. 10,000 ±0.5 ≤0.2