6 – LABNET BIOPETTE PLUS PIPETTORS
Nominal
Volume
(μL)
Volume
(μL)
Accuracy
(%)
Precision
(%)
Non-lter
Tips (μL)
8-channel and 12-channel Pipettors
8 - 10
12 - 10
Min. 0.5 ±10.0 ≤8.0
10
1 ±8.0 ≤6.0
5 ±4.0 ≤2.0
Max. 10 ±2.0 ≤1.2
8 - 50
12 - 50
Min. 5 ±4.0 ≤2.5
20025 ±3.0 ≤1.2
Max. 50 ±1.6 ≤0.6
8 - 200
12 - 200
Min. 20 ±3.0 ≤1.5
200100 ±1.5 ≤0.8
Max. 200 ±1.0 ≤0.6
8 - 300
12 - 300
Min. 50 ±1.6 ≤1.5
300150 ±1.2 ≤1.0
Max. 300 ±1.0 ≤0.6
The accuracy and precision were obtained gravimetrically
using manufacturer tips performing at least 10measurements
ofdistilled water at a temperature of 20°C ± 1°C according to
ENISO 8655 standards.
The use of tips from other manufacturers or lter tips may
result in incorrect liquid aspiration and require pipettor
recalibration.
The design of the pipettor enables the user to recalibrate it
according to the information presented in Section 9.