16 – LABNET BIOPETTE PLUS PIPETTORS
Checking the Pipetting Accuracy Parameters
Set the preset volume depending on the pipettor
volume according to the following table.
Nominal
volume
(µL)
Preset
volume
(μL)
Permissible
values (μL)
Volume Change ΔV (μL) for a
Turn of a Calibration Key by
1 Turn 1 Increment
Single-channel Pipettors
2 0.2 0.176 - 0.224 0.06 0.0025
10 1 0.975 - 1.025 0.33 0.0137
20 2 1.94 - 2.06 0.63 0.0262
100 10 9.84 - 10.16 2.50 0.104
200 20 19.76 - 20.24 6.30 0.262
1,000 100 98.4 - 101.6 25.00 1.04
5,000 500 494 - 506 125.00 5.2
10,000 1,000 975 - 1,025 250.00 10.4
8-channel and 12-channel Pipettors
10 1 0.92 - 1.08 0.33 0.0137
50 5 4.8 - 5.2 1.67 0.070
200 20 19.4 - 20.6 6.30 0.262
300 50 49.2 - 50.8 10.00 0.42
Perform 10 aspirations, and calculate the average value
in [mg].
Calculate the volume in [μL] by multiplying the value
in[mg] by the conversion factor Z [μL/mg].
If the average aspirated volume exceeds the permissible
values of the range, the pipettor should be recalibrated.