EasyManua.ls Logo

Lakeshore DRC-84C - Page 56

Default Icon
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DIN Standard Curve
T(Set Point) Data (Pin 1-4, 5-8, 9-12, 13-16)
077.0
0111
1010
1101
1100
130.0
0111
0101
0001
1111
300.0
0110
0011 1011 1011
The previous sets of data were given for the purpose of data veri-
fication.
If
the data does not match, refer to Section
6.5
to find the
problem.
b) D/A Calibrate
The calibration of
the D/A converter is an iterative process
using a span and offset potentiometer to get a calibrated
reading.
The reading can be calibrated in one of two methods:
(1) Method
1
-
Use of the Voltmeter
With the voltmeter connected between ground and TP#1,
a negative voltage (for diodes) should be present that
corresponds to the set point temperature from the front
panel thumbwheel switches.
As
an example, Curve D is
used to calibrate the instrument for a D-curve sensor;
the curve the instrument has can be noted from the sticker
on PROM U20. The curve is denoted in parenthesis
(*).
A temperature is dialed in to give a fairly low voltage
reading, for example 330.0K. The voltage read on the
voltmeter should be -.29111 V. Trimpot
R41
is varied
until this reading is obtained. Then a temperature
point is selected to give a higher voltage reading,
for example, 21.0K. The voltage read on the voltmeter
should be
-1.3505
V. Trimpot
R42
is varied until this
reading is obtained. This process should be repeated
until no further trim adjustments are needed.
(2) Method 2
-
Use of Display
The DRC-84C incorporates a method by which the D/A con-
verter can be tied to the input sensor voltage line.
With no sensor in either rear panel connector, TP1 and
TP2 should be tied together. This links the D/A output
with the A/D converter input.
The same iterative process should be used
as
in Method
(1)
with the exception that the display should read the
setpoint
setting
.
-50-

Table of Contents