Integral Process Thermostats and High-Temperature Thermostats
Operation manual
IN 130 T, IN 230 T, IN 230 TW, IN 530 T, IN 530 TW, IN 1030 T, IN 1330 TW, IN 1830 TW, IN 150 XT,
IN 250 XTW, IN 280 XT, IN 280 XTW, IN 550 XT, IN 550 XTW, IN 590 XTW, IN 750 XT, IN 950 XTW, IN 1590 XTW,
IN 1850 XTW, IN 2560 XTW, IN 2050 PW, IN 2560 PW, IN 4 XTW, IN 8 XTW
V6R9
Read this manual prior to performing any task!

Table of Contents

Questions and Answers

  • R
    Rachel MarquezSep 9, 2025
    Why is degassing functioning poorly on my Lauda Thermostat?
    • L
      Larry ChambersSep 9, 2025
      Degassing on your Lauda Thermostat might function poorly if pressure control is active, so switch it off. A pump level that is too high can also cause this, so select a lower pump level. Reduce the heating output, as it may be too high. If the cooling unit is active, switch it off. If the heat transfer liquid is heavily contaminated, change it. Ensure the filler nozzle is not closed by removing the tank lid and placing it loosely on the filler nozzle. If applicable, connect the inert gas blanket, with compressed air for pure degassing if necessary.
  • N
    Nancy HarringtonSep 12, 2025
    What does 'Pump blocked' mean on Lauda Thermostat display?
    • C
      Christine LeeSep 12, 2025
      The 'Pump blocked' message on your Lauda Thermostat display indicates a pump motor monitoring issue, such as overload or blockage. The viscosity of the heat transfer liquid may be too high; consider changing to another heat transfer liquid or increasing the set temperature.
  • J
    Jeanne HayesSep 13, 2025
    What does 'Evaporator frozen' mean on Lauda Thermostat display?
    • B
      Brian MossSep 14, 2025
      The 'Evaporator frozen' message on your Lauda Thermostat display indicates that the evaporator is frozen. Increase the pump level. Open the manual bypass in the constant temperature equipment. Increase the flow rate through the application.
  • A
    Albert WebbSep 16, 2025
    What does 'Low level (pump)' mean on Lauda Thermostat display?
    • L
      Lisa ThompsonSep 16, 2025
      The 'Low level (pump)' message on your Lauda Thermostat display indicates that the pump motor is idling. Check whether the float in the expansion vessel is blocked by a foreign object.

Summary

Safety

Safety information

Carefully read the operating manual before using the device. Keep it handy and save for future reference.

Unpacking

Electric shock

Instructions to prevent electric shock during unpacking and commissioning.

Commissioning

Setting heat transfer liquid

Procedure for selecting and configuring the heat transfer liquid used in the device.

Operation

General warning

Key safety warnings related to heat transfer liquid and device operation.

Setting Tmax

Procedure for setting the maximum temperature switch-off point for safety.

Maintenance

Warning notes for maintenance

Essential safety precautions before performing any maintenance tasks.

Faults

Alarms, errors and warnings

How the device displays and signals alarms, errors, and warnings.

Troubleshooting

Guide to identify and resolve common faults and issues with the device.