Integral Process Thermostats and High-Temperature Thermostats
Operation manual
IN 130 T, IN 230 T, IN 230 TW, IN 530 T, IN 530 TW, IN 1030 T, IN 1330 TW, IN 1830 TW, IN 150 XT,
IN 250 XTW, IN 280 XT, IN 280 XTW, IN 550 XT, IN 550 XTW, IN 590 XTW, IN 750 XT, IN 950 XTW, IN 1590 XTW,
IN 1850 XTW, IN 2560 XTW, IN 2050 PW, IN 2560 PW, IN 4 XTW, IN 8 XTW
V6R9
Read this manual prior to performing any task!

Table of Contents

Questions and Answers

  • C
    Charles BrownSep 10, 2025
    Why is degassing functioning poorly on my Lauda Thermostat?
    • B
      Brandon MillsSep 10, 2025
      Degassing on your Lauda Thermostat might function poorly if pressure control is active, so switch it off. A pump level that is too high can also cause this, so select a lower pump level. Reduce the heating output, as it may be too high. If the cooling unit is active, switch it off. If the heat transfer liquid is heavily contaminated, change it. Ensure the filler nozzle is not closed by removing the tank lid and placing it loosely on the filler nozzle. If applicable, connect the inert gas blanket, with compressed air for pure degassing if necessary.
  • A
    Andrew HoodSep 12, 2025
    What does 'Pump blocked' mean on Lauda Thermostat display?
    • S
      sfarrellSep 12, 2025
      The 'Pump blocked' message on your Lauda Thermostat display indicates a pump motor monitoring issue, such as overload or blockage. The viscosity of the heat transfer liquid may be too high; consider changing to another heat transfer liquid or increasing the set temperature.
  • M
    Mark JonesSep 14, 2025
    What does 'Evaporator frozen' mean on Lauda Thermostat display?
    • M
      Michelle NolanSep 14, 2025
      The 'Evaporator frozen' message on your Lauda Thermostat display indicates that the evaporator is frozen. Increase the pump level. Open the manual bypass in the constant temperature equipment. Increase the flow rate through the application.
  • A
    Adam ReyesSep 16, 2025
    What does 'Low level (pump)' mean on Lauda Thermostat display?
    • B
      Brandon HillSep 16, 2025
      The 'Low level (pump)' message on your Lauda Thermostat display indicates that the pump motor is idling. Check whether the float in the expansion vessel is blocked by a foreign object.

Summary

1 Safety

1.1 Safety Information

General safety information, warnings, and safe state definitions for device operation.

1.2 Operator Obligations

Operator responsibilities, regulatory compliance, and requirements for system installation.

1.15 Safety Fittings

Details overtemperature protection and low-level protection safety units on the device.

2 Unpacking

2 Unpacking Procedures

Instructions for unpacking the device, checking for completeness, and handling transport damage.

3 Device Description

3.4 Integral Large casing with pressure overlay

Visual overview of large casing Integral devices with pressure overlay features.

3.7 Hydraulic Circuit

Explanation of the hydraulic circuit in Integral T and Integral XT devices.

3.8 Rating Label

Details on the LAUDA device rating label specifications and their meaning.

3.9 Interfaces

Overview of standard and accessory interface modules for device connectivity.

4 Before Starting Up

4.1 Install Device

Guidelines for correct device installation, handling, and environmental conditions.

4.3 Connecting an External Application

Instructions for connecting external applications, including safety warnings.

4.4 LAUDA Heat Transfer Liquids

Information on permissible heat transfer liquids, their properties, and handling guidelines.

5 Commissioning

5.1 Establishing a Mains Connection

Procedure for connecting the device to the mains power supply.

5.3 Operating the Device with Operating Unit

Guide to operating the device using the control unit, including navigation and windows.

5.4 Setting Heat Transfer Liquid

Instructions for selecting and setting the heat transfer liquid used in the device.

5.5 Filling the Device

Procedure for filling the device with heat transfer liquid, including safety warnings.

5.6 Degassing Heat Transfer Liquid

Steps for degassing the heat transfer liquid using the low boiling point expulsion method.

6 Operation

6.1 General Warning

General safety warnings for device operation, including hazards related to heat transfer liquid.

6.23 Safe Mode Safety Fittings

Configuration and activation of Safe Mode for emergency device operation.

8 Operating a Device with Pressure Overlay

8.1 Warnings and Safety Instructions

Safety guidelines and warnings specific to devices with pressure overlay.

9 Maintenance

9.1 Warning Notes for Maintenance

Safety precautions and protective equipment required before performing maintenance.

10 Faults

10.1 Alarms, Errors, and Warnings

Explanation of how alarms, errors, and warnings are displayed and handled.

10.3 Troubleshooting

Table of common faults, their causes, and possible remedies for process thermostats.

Lauda IN 2560 XTW Specifications

General IconGeneral
Temperature stability±0.01 °C
Power consumption2.7 kW
Power supply230V 1~ 50/60Hz
Cooling capacity at 20°C0.55 kW
Cooling capacity2560 W at 20°C
InterfaceRS232