OPERATION MANUAL
ultracool mini
0140 Indigo
Ref: MI-080 Rev: 1 Date: 30.04.03
50/60 Hz

Questions and Answers

  • D
    David ThompsonAug 3, 2025
    Why is my Lauda Chiller alarming due to high refrigerant pressure?
    • B
      Brian VasquezAug 3, 2025
      The Lauda Chiller may alarm due to high refrigerant pressure for several reasons. The lateral panels of the housing might be open, so close them. There may be low airflow into the condenser, so clean the condenser. The ambient temperature could be too high, in which case you should wait until it lowers. The water temperature might be too high; try to cool down the water in the circuit while the chiller is stopped. Finally, the motor fan may not be working; check the motor fan fuses, and if the problem persists, contact authorised technical service.
  • A
    Anthony ThorntonAug 8, 2025
    What to do if the Lauda Ultracool Superplus Chiller displays LO HI E0, Er2, EE codes?
    • L
      Laura ReevesAug 9, 2025
      If your Lauda Chiller's control thermostat displays LO HI E0, Er2, or EE codes, it could mean the temperature has been below the setpoint for some time. If the compressor is stopped, the water temperature will rise to normal eventually. It could also mean the temperature has been above the setpoint for some time; if the compressor is still running, contact authorised technical service. The thermostat or temperature sensor might also be faulty; check that the cold water setpoint is within the indicated limits, and if the problem persists, contact authorised technical service.
  • K
    Kyle WilliamsAug 14, 2025
    What to do if Lauda Ultracool Superplus's low flow/temperature control operates continuously?
    • A
      Ann WoodAug 14, 2025
      If the low flow/temperature control of your Lauda Chiller operates continuously, the water circuit may be blocked. Clean the water circuit and check for closed valves. Freezing could also be a factor, so check the proportion of ethylene glycol. Another possibility is a circuit break in the pump; reset the circuit breaker of the pump, and if the problem persists, contact authorised technical service.
  • M
    Michael JonesAug 21, 2025
    What to do if there is a water level alarm on my Lauda Chiller?
    • M
      Maria AdamsAug 21, 2025
      If your Lauda Chiller is showing a water level alarm, it is likely due to water leakage. Check the water circuit and refill the tank.
  • J
    Jonathan EdwardsAug 27, 2025
    How to fix a Lauda Ultracool Superplus with an alarm due to low refrigerant pressure?
    • K
      Karen MercerAug 27, 2025
      If your Lauda Chiller is alarming due to low refrigerant pressure, this could be caused by a gas leakage. Contact authorised technical service.

Summary

General Safety Warnings

Introduction to Ultracool Unit

Safety Regulations and Best Practices

Outlines essential safety regulations for operation, maintenance, and repair, including using correct tools and parts.

Unit Installation Guide

Water and Electrical Connections

Guidance on water filter installation, pipe sizing, thermal insulation, and electrical supply requirements.

Unit Start-up Procedures

First Time Chiller Start-up

Step-by-step guide for powering on, heating the crankcase, filling the tank, and checking pump rotation.

Control Panel Overview

Control Thermostat Operation and Alarms

Instructions for setting temperature, silencing alarms, and interpreting thermostat codes.

Routine Maintenance Guide

Troubleshooting Common Issues

Alarm and Default Resolution

Guide to diagnosing and resolving common alarms like high/low refrigerant pressure and water level issues.

Unit Technical Specifications

Annexes and Water Quality

Spare Parts Identification

Electrical Diagram

Lauda Ultracool Superplus series Specifications

General IconGeneral
BrandLauda
ModelUltracool Superplus series
CategoryChiller
LanguageEnglish