INSTRUCTIONS MANUAL 
-20- 
Refer to collapsible tracer cartridge precautions [MANU008.ATS]. 
EMPTYING THE USED OIL CONTAINER 
This operation must be performed whenever the oil level exceeds 200 cc. Procedure: remove 
the container from its lodging very carefully and without exerting pressure on the scale. 
Unscrew the container while holding the cap in place; empty the used oil into a suitable 
container for used oils. Screw the container back on while holding the cap in place. Carefully 
replace the container in its lodging  without exerting pressure on the scale. N.B.: In order to 
avoid damage to the oil scale, never exert pressure on it either from above or from below.  
REPLACING THE PRINTER PAPER 
Use only heat-sensitive paper of the type described below. 
Paper width: 58 mm  
Maximum paper roll diameter: 40mm   
UNSCHEDULED MAINTENANCE / CALIBRATION 
Perform this operation whenever the values displayed on the LCD do not correspond to real 
values. WARNING: The operations listed below must be performed with maximum attention 
and care. In particular, always observe the following precautions. Always place the weights at 
the center of the scale plate. Never exert pressure on the oil scale. Always recover the gas in 
the high- and low-pressure hoses before beginning calibration of the transducers.  
To enter the CALIBRATION MENU: 
Switch the machine on. The MAIN MENU will appear after a few seconds: 
     
Select NEXT MENU: 
     
Now select CALIBRATION from the menu:  
   
Type the code 0791 and press ENTER to confirm. 
CALIBRATING THE BOTTLE SCALE 
Disconnect the machine from the electrical supply. Procure a known reference weight from 16 
to 18 kg. Remove the plastic cover on the rear of the machine to access the machine bottle. 
Close the blue and red taps on the bottle. Unscrew the bottle lock nut (3 – Fig. 4). Separate the 
heating coil (r – Fig. 8) from the bottle (do not touch or disconnect the wires of the resistance 
coil). Remove the bottle (d – Fig. 8) from its lodging, leaving the resistance around the scale 
plate. Set the bottle on a stand at least 40 cm in height. 
Access the CALIBRATION MENU as explained above.  
The following will be displayed: 
To enter calibration 
insert code 
.... 
And press ENTER        
AUTOMATIC PROCEDURE  
ASSISTED PROCEDURE   
NEXT MENU            
Gas Avail  xxxxx G 
 
<CALIBRATION>          
DATA AND CONFIGURAT. 
SERVICE ALARM 
BOTTLE FILLING