What to do if Leica Rugby Measuring Instruments does not turn on?
- MMatthew HudsonAug 2, 2025
If your Rugby won't turn on, the batteries may be low or dead. Check the batteries, and replace or charge them if needed.
What to do if Leica Rugby Measuring Instruments does not turn on?
If your Rugby won't turn on, the batteries may be low or dead. Check the batteries, and replace or charge them if needed.
What to do if Leica Rugby 840 laser receiver is not working properly?
If the laser receiver isn't working correctly, there could be a few reasons: * The Rugby might not be rotating, possibly because it's levelling or in Elevation Alert. Check the Rugby's operation. * The receiver could be out of range. Try moving closer to the Rugby. * The receiver batteries might be low. Replace the receiver batteries.
Why Leica Rugby Measuring Instruments is working, but not self-levelling?
If your Rugby isn't self-levelling, it's likely because it's in Manual Mode. To enable self-levelling, switch the Rugby to Automatic Mode by pressing the Automatic/Manual Mode button. When in Automatic Mode, the X-axis and Y-axis LEDs will flash green while levelling. In Manual Mode, these LEDs will be red.
How to enable Elevation Alert function on Leica Measuring Instruments?
To enable or disable the Elevation Alert function: Turn on the Rugby so it is rotating, then press and hold the Left and Right Arrow buttons simultaneously. While holding these buttons, press the Automatic/Manual Mode button. The Rugby will beep once to confirm the change.
What to do if the distance of the laser is reduced on Leica Measuring Instruments?
If the laser distance is reduced, dirt might be reducing the laser output. Clean the windows of both the Rugby and the receiver.
How to fix Leica Rugby Measuring Instruments that turns on with the last mode saved?
If your Rugby turns on with the last mode saved, this function can be disabled. With the Rugby turned on and rotating, press the power button to turn the Rugby off. Then, press and hold both the Automatic/Manual Mode button and the power button for five seconds. The Rugby will beep once to indicate the change.
Why Leica Rugby Measuring Instruments does not turn on in Automatic Mode?
The Rugby is designed to always turn on in Automatic Mode unless specifically disabled by the user. The Automatic Mode can be enabled or disabled by pressing the Automatic/Manual Mode button.