Leica Stellaris 8 FALCON quick manual
Version 1.0
Question and Answer IconNeed help?

Do you have a question about the Leica Stellaris 8 FALCON and is the answer not in the manual?

Leica Stellaris 8 FALCON Specifications

General IconGeneral
Microscope TypeConfocal
Objective TurretMotorized
SoftwareLAS X
FALCONFast Lifetime Contrast
Imaging ModeFLIM, confocal
Laser Lines470-670 nm
Spectral DetectionYes, spectral detection with freely adjustable bandwidth
DetectorsPMT
Detector TechnologyGaAsP and PMT
Scanner TypeResonant and Galvo
ApplicationLive cell imaging, FRAP, FRET
DimensionsVaries with configuration
WeightVaries with configuration

Summary

Microscope Start-up Procedure

Stage Adapter Check and Setup

Verify the installed stage adapter and replace if necessary, noting specific types for wellplates and dishes.

System Power On Sequence

Instructions for powering on the microscope, laser, and fluorescent lamp using the main unit's buttons.

Software and Configuration Setup

Computer Login and Software Launch

Start the computer, log in with credentials, and launch the Las X software application.

LAS X Configuration Selection

Choose the appropriate configuration within the Leica Application Suite X, typically 'machine.xlhw'.

Sample Insertion and Laser Configuration

Insert the sample after software startup and select the desired laser configuration.

Excitation and Detection Setup

Laser Activation for Sample Dyes

Turn on the Diode 405 laser for DAPI/Hoechst and the white light laser (WLL) if applicable.

Acquisition Paradigm Configuration

Configure excitation/detection by selecting saved schemes or creating new ones using dyes and detectors.

Detector Usage and Heating/CO2 Control

HyD Detector Selection Guide

Guidance on using HyD S for bright channels and HyD X for dim channels or FLIM applications.

Environmental Controller Operation

Instructions for operating the environmental controller, including heating and CO2 settings.

Microscope Shutdown Procedures

Pre-Shutdown Checks and Protocols

Save data, check next user's needs, and follow protocol A (short-term) or B (long-term) shutdown.

Protocol A: Short-Term Shutdown

Turn off specific lasers, shut down software, clean objectives, log off, and manage heating/CO2.

Protocol B: Long-Term Shutdown

Turn off all lasers, shut down software, clean objectives, microscope, and fluorescent lamp.