Do you have a question about the LEYBOLD THERMOVAC TTR 91 R and is the answer not in the manual?
Supply Voltage | 15 to 30 V DC |
---|---|
Output Signal | 0 to 10 V DC |
Process Connection | DN 16 ISO-KF |
Housing Material | Stainless steel |
Sensor Type | Pirani |
Accuracy | +/- 15% of reading |
Measurement Principle | Thermal conductivity |
Defines safety symbols and their meanings within the manual.
Specifies the required technical training and experience for users.
Provides overarching safety guidelines and precautions for installation and operation.
Details manufacturer's liability and warranty conditions, voiding clauses.
Lists detailed specifications, dimensions, electrical, and material properties.
Enlists compatible accessories, cables, and spare parts with their part numbers.
Outlines the availability and process of calibration services for Leybold gauges.
Provides specific wiring diagrams and pinout details for optimal electrical setup.
Explains pressure reading, output voltage scaling, and conversion formulae.
Details how different gases affect pressure readings and provides correction factors.
Guides on configuring and using the set-point output for process control.
Covers error detection, status LED indications, and error voltage outputs.
Instructions for preparing the device for high-temperature bakeout to achieve lower base pressure.
Explains drift and the need for recalibrating atmosphere and vacuum settings.
Step-by-step guide for calibrating the gauge at atmospheric pressure.
Steps for adjusting the gauge to read correctly at low vacuum pressures.
How to perform atmosphere and vacuum adjustments using an external switch.
Procedure for recalibrating the gauge after replacing the sensor tube.
Instructions for safely removing and fitting a replacement sensor tube.