FC5 HORNET
OPERATIONS MANUAL
Question and Answer IconNeed help?

Do you have a question about the Lifeloc FC5 HORNET and is the answer not in the manual?

Summary

FC5 Hornet Components Overview

FC5 Hornet Battery Installation

FC5 Hornet Precautions and Guidelines

Pre-Test Procedures

Wait 10 minutes after drinking, smoking, or oral intake before testing.

Device Integrity Warnings

Avoid blowing smoke into the port and use authorized technicians for service.

FC5 Hornet Test Modes Explained

POS+;- NEG Test Mode

Displays results as POS+ or NEG for zero tolerance environments.

LEVELS Test Mode

Alerts operator to alcohol levels with manual override for difficult subjects.

OPEN CONTAINER Test Mode

Samples air-space over open containers for precise testing.

FC5 Hornet Operations

Getting Started

Instructions on turning on the FC5, self-test, and selecting test modes.

Conducting an Automatic Test

Steps to position the cone, instruct the subject, and read results.

Conducting a Manual Test

Steps for performing a manual breath test using the Execute button.

Conducting an Open Container Test

Procedure for sampling air-space over open containers or spills.

Conducting an Ambient Air Test

How to test ambient air by waving the unit and pressing Execute.

FC5 Hornet User Prompts

FC5 Hornet Specifications and Calibration

Specifications Details

Key technical specifications including sensor, power, size, and warranty.

Calibration Requirements

Details on periodic calibration by a certified technician every 12 months.

FC5 Hornet Service and Maintenance

General Cleaning

Wipe unit with a soft cloth and non-abrasive disinfectant cleaner.

Fuel Cell Longevity

Perform 2-3 tests every 30 days to prevent fuel cell drying.

FC5 Hornet Product Warranty

General Warranty Terms

Warrants equipment free from defects for 12 months from original invoice date.

Warranty Exclusions

Warranty void if altered, repaired by unauthorized personnel, or uses non-approved parts.

Limitation of Liability

Excludes liability for indirect, incidental, or consequential damages.

FC5 Hornet Fuel Cell Warranty

Fuel Cell Coverage

Warrants fuel cell operates without defect for the life of the instrument.

Warranty Conditions

Conditions include proper use, maintenance, and original purchaser status.

Warranty Invalidation

Warranty invalidated by misuse, abuse, modification, or defacement of serial number.

Overview

The Lifeloc FC5 Hornet is a new generation of alcohol screener designed for use in various safety-sensitive industries, including schools, police, probation, jails, and mining. This device simplifies alcohol testing by eliminating the need for mouthpieces or special attachments, making it user-friendly and efficient. Its design incorporates icons to overcome language barriers, ensuring clear communication of test results.

At its core, the FC5 Hornet is alcohol-specific, utilizing a platinum fuel cell that provides rapid response and reset times, contributing to a long working life for the device. A key safety feature is its ability to allow the operator to view test results without revealing the display to the subject, maintaining discretion and privacy.

The device offers three primary operating test modes: Subject, Ambient Air, and Open Container testing.

Function Description

The FC5 Hornet is a portable breath alcohol tester that measures the presence and levels of alcohol. It is designed for quick and easy operation, providing reliable results for screening purposes.

Usage Features

Getting Started:

To begin, the user simply presses the Power button momentarily. The FC5 Hornet will initiate a self-test and then display the POS/NEG test mode as the default. The device is designed to automatically power off after 5 minutes of inactivity to conserve battery life.

Test Modes:

The FC5 Hornet offers three distinct test modes, which can be scrolled through using the Function button:

  1. POS+/- NEG Test Mode: This mode is ideal for zero-tolerance environments. Results are displayed as either "POS +" (any alcohol level > .000 BAC) or "NEG" (no alcohol detected). Tests are automatically triggered when the subject exhales sufficient breath towards the sampling cone. For uncooperative subjects or discrete ambient air testing, a manual override is available via the Execute button.

  2. LEVELS Test Mode: For more precise alcohol testing, this mode alerts the operator to specific levels of alcohol. It features automatic testing with manual override capabilities for challenging subjects. The pre-set alcohol levels are categorized as:

    • Zero: <.010% BAC (<.04 mg/l)
    • Low: .011 - .020% BAC (.05 - .09 mg/l)
    • High: >.021% BAC (>.100 mg/l) The display uses distinct icons to represent these levels, ensuring clear interpretation.
  3. OPEN CONTAINER Test Mode: This mode is designed to sample the air-space over open drink containers or spills. It employs a unique double pump action to ensure a sufficient test sample, even from small-neck bottles. One-button sample triggering allows for precise positioning over the sample, and the quiet pump action ensures discrete and unobtrusive sampling.

Conducting Tests:

  • Automatic Test:

    1. Position the sampling cone approximately 4 inches (10 cm) from the subject's mouth.
    2. Instruct the subject to blow steadily towards the cone.
    3. The test will automatically trigger once an adequate sample is provided.
    4. The results will be displayed and remain on screen for 30 seconds or until the Function button is pressed.
  • Manual Test:

    1. Position the sampling cone no less than 4 inches from the subject's mouth.
    2. Instruct the subject to blow steadily towards the cone.
    3. While the subject is blowing, press the Execute button to manually trigger the test.
    4. Results will be displayed for 30 seconds or until the Function button is pressed.
  • Open Container Test:

    1. Position the sampling cone no less than 4 inches from the container or sample.
    2. Press the Execute button to trigger the test.
    3. For optimal results, it is recommended to slightly agitate the sample prior to triggering the test.
    4. Results will be displayed for 30 seconds or until the Function button is pressed.
  • Ambient Air Test:

    1. In any of the three test modes, wave the unit in the air.
    2. Press the Execute button to trigger the test.
    3. Results will be displayed for 30 seconds or until the Function button is pressed.

User Prompts:

The FC5 Hornet uses clear icons and text prompts to guide the user:

  • "POS+/-NEG" indicates the POS/NEG Test Mode, where NEG results register as Zero and POS results are anything >0.
  • "- NEG" signifies negative results in POS/NEG Test Mode.
  • "POS +" signifies positive results in POS/NEG Test Mode.
  • A blowing icon with "BLOW" indicates the Levels Test Mode, showing results across Zero, Low, and High alcohol content.
  • Specific icons represent <.010% BAC, .011 - .020% BAC, and >.021% BAC in Levels Test Mode.
  • An icon of a mug with "+/- GO" indicates Open Container Test Mode, where NEG results register as Zero and POS results are anything >0.
  • A mug with a minus sign indicates negative results in Open Container Test Mode.
  • A mug with a plus sign indicates positive results in Open Container Test Mode.
  • An hourglass icon appears when the FC5 is in wait mode, which can occur while the device is recovering from a POS test.

Battery Installation:

The FC5 Hornet operates on 4 AA alkaline batteries. To install or change batteries, the protective grip must first be removed. The battery door is located on the lower back of the device and is removed by pressing in and down. Batteries should be installed according to the symbols in the battery case, and the door is closed by pushing up until it locks shut.

Precautions:

To ensure accurate results and prevent damage, users should observe the following precautions:

  • Wait at least 10 minutes after drinking, smoking, or inserting anything into the mouth before taking a breath test.
  • Avoid blowing smoke into the port, as this can damage the sensor.
  • The FC5 Hornet should be stored in temperatures between 32° - 130° F (0° - 55° C).
  • There are no user-serviceable parts inside the device; all servicing must be performed by an authorized FC5 technician.

Maintenance Features

Calibration:

To maintain accuracy, all breath testers, including the FC5 Hornet, require periodic calibration. It is recommended that the device be calibrated by a factory-certified calibration technician at least once every 12 months. For calibration or service, users can contact Lifeloc via email at service@lifeloc.com, visit their website at www.lifeloc.com, call 1.800.722.4872, or reach out to their local distributor.

General Maintenance:

For general upkeep, the unit should be wiped with a soft cloth sprayed with a non-abrasive disinfectant cleaner. To ensure the longest possible life for the FC5 Hornet, it is advised to perform two or three tests at least every 30 days. This practice helps prevent the fuel cell from drying out during extended periods of non-use. If service is required, the same contact methods as for calibration apply.

Warranty:

Lifeloc Technologies, Inc. ("Lifeloc") warrants that new Lifeloc Equipment, including the FC5 Hornet, purchased through Lifeloc or authorized distributors, will be free from defects in material and workmanship under normal use and service for twelve (12) months from the original invoice date. This warranty covers all portable hand-held and fixed station breathalyzers manufactured by Lifeloc, as well as associated printers, keyboards, cables, cases, and power supplies. Software is warranted to perform substantially in accordance with Lifeloc user manuals and to be free from defects for twelve (12) months from the original invoice date.

The warranty does not apply if the product has been altered, customized, repaired, reported stolen, or modified by unauthorized personnel, or if non-Lifeloc approved parts are used for replacement or repair. Lifeloc's exclusive liability for a breach of warranty is limited to repair, replacement, or a refund of the original purchase price. Lifeloc is not responsible for customer software, settings, or test records on returned products. The warranty is non-transferable and applies to equipment purchased from and after September 1, 2014.

Fuel Cell Warranty:

The fuel cell purchased with the instrument is warranted to operate without defect for the life of the instrument, subject to specific conditions:

  1. Lifeloc may choose to repair or replace any defective fuel cell.
  2. Defective fuel cells returned to Lifeloc become their property.
  3. This warranty applies only to fuel cells purchased with new instruments after June 1, 2014.
  4. The warranty is valid only for the original purchaser and does not transfer if the instrument is resold, stolen, donated, or given to another party.
  5. The warranty is invalidated if the instrument has been misused, abused, modified, or improperly maintained. Lifeloc will conduct a billable Factory Diagnostic Test on all instruments submitted for warranty evaluation.
  6. The warranty does not apply if the product has been repaired or modified by unauthorized technicians or if non-Lifeloc approved parts are used.
  7. Upon subsequent instrument trade-in, no residual monetary value is associated with the fuel cell.
  8. The Lifeloc Limited Fuel Cell Warranty does not extend beyond the end-of-service and support life of the instrument as determined by Lifeloc Technologies.
  9. All Lifeloc fuel cells are otherwise subject to Lifeloc's standard Warranty.

The lifetime fuel cell warranty is valid throughout the US and Canada. International customers should contact Lifeloc directly for specific terms and conditions.

Lifeloc FC5 HORNET Specifications

General IconGeneral
Sensor TypeFuel Cell
Resolution0.001 % BAC
Warranty1 year limited warranty
TypeBreathalyzer
DisplayLCD
Response TimeWithin seconds
Operating Temperature32°F to 122°F (0°C to 50°C)