Do you have a question about the Lifeloc FC20 and is the answer not in the manual?
Measurement Range | 0.000 to 0.400% BAC |
---|---|
Resolution | 0.001% BAC |
Response Time | Less than 10 seconds |
Display | LCD |
Warranty | 1 year |
Type | Breath Alcohol Tester |
Sensor Type | Fuel Cell |
Accuracy | ±0.005% BAC at 0.100% BAC |
Warm-up Time | 10 seconds |
Power Supply | 4 x AA batteries |
Battery Life | Approximately 500 tests |
Storage Temperature | -20°C to 60°C |
Humidity Range | 5% to 95% non-condensing |
Calibration | Recommended every 6-12 months |
Step-by-step instructions for installing 4 AA alkaline batteries into the FC20 unit.
Explains the correct procedure for attaching a mouthpiece to the FC20 for breath testing.
Details the process for powering the FC20 unit on and off using its power button.
Advises on the 15-minute observation period before testing to prevent mouth alcohol contamination.
Guides the user through performing an automatic breath alcohol test using the FC20.
Guides on printing current, specific, or all stored breath test results, covering various print options.
Provides step-by-step instructions for performing an automatic breath alcohol test.
Details the procedure for conducting a manual breath alcohol test when automatic testing is not feasible.
Outlines the method for performing a passive breath alcohol screen without a mouthpiece.
Explains the purpose of calibration and Lifeloc's recommendations for calibration frequency.
Guides on setting the specific BAC level or corrected standard for calibration.
Instructions for preparing the wet bath simulator for calibration, including filling and heating.
Step-by-step guide for performing a wet bath calibration using the FC20 and simulator.
Details the procedure for performing a wet bath calibration check and expected results.
Instructions for setting up the dry gas cylinder and regulator for calibration.
Step-by-step guide for performing a dry gas calibration using the FC20.
Details the procedure for performing a dry gas calibration check and expected results.
Lists common FC20 error messages and provides corresponding corrective actions.