Calibration Settings
SETTING THE CALIBRATION STANDARD
From the “CALIBRATION” display, press the Exe-
cute button. The display now reads “XXX CHECK”.
Press Function button until the display reads “CAL
SETTINGS“.
Press the Execute button and the display will read
“Cal Standard”
Use the + and – buttons to change the number to
the BAC level of standard you will be calibrating
to (it should be the same as on the bottle of certi-
fied solution or your corrected standard if using a
dry gas tank).
Press the Function button to move to the next set-
ting or repeatedly until the display reads Exit.
Press the Execute button to save the changes and
return to the testing mode.
Once you set the standard you do not have to set
it again unless you change to a different method
of performing a calibration/calibration check.
Cal Standard
0.100
Page 38
Calibration Settings
SETTING THE CALIBRATION STANDARD
From the “CALIBRATION” display, press the Exe-
cute button. The display now reads “XXX CHECK”.
Press Function button until the display reads “CAL
SETTINGS“.
Press the Execute button and the display will read
“Cal Standard”
Use the + and – buttons to change the number to
the BAC level of standard you will be calibrating
to (it should be the same as on the bottle of certi-
fied solution or your corrected standard if using a
dry gas tank).
Press the Function button to move to the next set-
ting or repeatedly until the display reads Exit.
Press the Execute button to save the changes and
return to the testing mode.
Once you set the standard you do not have to set
it again unless you change to a different method
of performing a calibration/calibration check.
Cal Standard
0.100