Page 40
Calibration Settings
SETTING THE CALIBRATION CHECK LOCK-OUT
From the “CAL SETTINGS” display, press the Exe-
cute button.
Press the Function button until the display reads
“Check Period XXX”.
Press the +/- buttons to enter the number of days
until you require the next cal check of your FC20.
Choosing “0” turns this setting off.
Press the Function button to move to the next set-
ting or repeatedly until the display reads “Exit”.
Press the Execute button to save the changes and
return to the testing mode.
SETTING THE CALIBRATION LOCK-OUT PERIOD
From the “CAL SETTINGS” display, press the Exe-
cute button.
Press the Function button until the display reads
“Cal Period XXX”.
Press the +/- buttons to enter the number of days
until you require the next calibration of your FC20.
Choosing “0” turns this setting off.
Press the Function button to move to the next set-
ting or repeatedly until the display reads “Exit”.
Press the Execute button to save the changes and
return to the testing mode.
Page 40
Calibration Settings
SETTING THE CALIBRATION CHECK LOCK-OUT
From the “CAL SETTINGS” display, press the Exe-
cute button.
Press the Function button until the display reads
“Check Period XXX”.
Press the +/- buttons to enter the number of days
until you require the next cal check of your FC20.
Choosing “0” turns this setting off.
Press the Function button to move to the next set-
ting or repeatedly until the display reads “Exit”.
Press the Execute button to save the changes and
return to the testing mode.
SETTING THE CALIBRATION LOCK-OUT PERIOD
From the “CAL SETTINGS” display, press the Exe-
cute button.
Press the Function button until the display reads
“Cal Period XXX”.
Press the +/- buttons to enter the number of days
until you require the next calibration of your FC20.
Choosing “0” turns this setting off.
Press the Function button to move to the next set-
ting or repeatedly until the display reads “Exit”.
Press the Execute button to save the changes and
return to the testing mode.